Page 1 - Quinton® Q-Stress® 4.5

Quinton® Q-Stress® 4.5USER’S GUIDE AND SERVICE MANUAL70-00755-01 D

Page 2 - QUINTON® Q-STRESS®

70-00755-01 D ContentsixContentsAppendix A: SpecificationsSystem Specifications ...

Page 3

Quinton® Q-Stress® 4.5 70-00755-01 D8-18 Final ReportsThis chapter describes the process of creating, reviewing, editing, and printing a final report

Page 4

70-00755-01 D Quinton® Q-Stress® 4.58-2Final ReportsEdit the Final Report Summary PageAfter you click the Recovery icon, the Edit Final Report icon b

Page 5 - Chapter 4: System Setup

Quinton® Q-Stress® 4.5 70-00755-01 D8-3Edit the Final Report Summary Pagefollowing figures show samples of preprogrammed statements for each category

Page 6 - 70-00755-01 D Contents

70-00755-01 D Quinton® Q-Stress® 4.58-4Final ReportsFast ReportsUse words or prepared phrases from the QuickLists utility to quickly annotate final r

Page 7 - Chapter 9: File Management

Quinton® Q-Stress® 4.5 70-00755-01 D8-5Review ModeReview ModeEntering Review Mode ends the display of waveform data. You cannot resume monitoring dat

Page 8 - Chapter 11: Options

70-00755-01 D Quinton® Q-Stress® 4.58-6Final ReportsView a Final ReportClick the View Final Report button to view a PDF file of your final report wit

Page 9 - Chapter 13: Maintenance

Quinton® Q-Stress® 4.5 70-00755-01 D8-7Review ModeSaving the TestTo leave Review Mode, press the Exit button. A confirmation dialog box appears (Figu

Page 10 - Appendix A: Specifications

70-00755-01 D Quinton® Q-Stress® 4.58-8Final ReportsXML (Q-Exchange Option Only)The Q-Stress network system provides the patient demographic, final r

Page 11 - Contents

Quinton® Q-Stress® 4.5 70-00755-01 D8-9Review ModeThe name of the default final report format is shown. To change to a different report format, click

Page 12 - 1 Safety

70-00755-01 D Quinton® Q-Stress® 4.58-10Final ReportsTabularSelect the template you want to use for your report. Use the scroll bar to see the entire

Page 13 - Indications for Use

Contents 70-00755-01 DxContents

Page 14 - Safety Definitions

Quinton® Q-Stress® 4.5 70-00755-01 D8-11Review ModePeak ExerciseSelect this section to print the peak exercise test values as part of the final repor

Page 15 - General Safety Statements

70-00755-01 D Quinton® Q-Stress® 4.58-12Final ReportsCustomizing a Trend Graph TemplateTo customize a trend graph template, select the custom templat

Page 16 - Symbols and Labels

Quinton® Q-Stress® 4.5 70-00755-01 D8-13Review ModeIn-Test ReportsTo include in-test reports (those waveform reports printed during the test except f

Page 17 - AHA to IEC Equivalents

70-00755-01 D Quinton® Q-Stress® 4.58-14Final ReportsSaving the Final Report FormatWhen you have made all your changes, click Save As. The Final Repo

Page 18 - 2 Quick Start

Quinton® Q-Stress® 4.5 70-00755-01 D9-19 File ManagementChapter contents◆ Database 9-2◆ Final Report Repository 9-3◆ PDF Files 9-4◆ Database Restore

Page 19 - Before the Test

70-00755-01 D Quinton® Q-Stress® 4.59-2File ManagementDatabaseThe data from each stress test is stored in the database that ships with the system. Q-

Page 20 - Rest Phase

Quinton® Q-Stress® 4.5 70-00755-01 D9-3Final Report RepositoryFinal Report RepositoryEvery time you print or view a final report, a PDF file is store

Page 21 - Exercise Phase

70-00755-01 D Quinton® Q-Stress® 4.59-4File ManagementPDF FilesQ-Stress saves both the final reports and the full disclosure files (Full Disclosure o

Page 22 - Review Phase

Quinton® Q-Stress® 4.5 70-00755-01 D9-5PDF FilesIf your Q-Stress system is equipped with a secondary hard disk (either external or internal) or if yo

Page 23 - Quick Start

70-00755-01 D Quinton® Q-Stress® 4.59-6File Management4. Use Windows Explorer to delete the patient folder and contents from the PDF repository.Resto

Page 24 - 3 System Overview

Quinton® Q-Stress® 4.5 70-00755-01 D1-11 SafetyChapter contents◆ Indications for Use 1-2◆ Safety Definitions 1-3◆ General Safety Statements 1-4◆ Symb

Page 25 - System Overview

Quinton® Q-Stress® 4.5 70-00755-01 D9-7Database BackupDatabase BackupYou cannot use the Q-Stress system while the backup is in process.Note: The back

Page 26 - Hardware

70-00755-01 D Quinton® Q-Stress® 4.59-8File Management8. Click Start to initiate the backup.9. Click Yes to acknowledge the time it will take to do t

Page 27 - Port Use

Quinton® Q-Stress® 4.5 70-00755-01 D9-9Database RestoreDatabase RestoreYou cannot use the Q-Stress system while the restore is in process.Note: The r

Page 28 - Isolation Transformer

70-00755-01 D Quinton® Q-Stress® 4.59-10File ManagementPurge the Database!WARNING. System unavailable.Purging data removes information permanently. I

Page 29

Quinton® Q-Stress® 4.5 70-00755-01 D9-11Purge the Database3. Click the Patient/Test(s) Deletion tab.4. Select one of the three Purge options:• Identi

Page 30 - Keyboard

70-00755-01 D Quinton® Q-Stress® 4.59-12File Management9. The records are deleted and space in the database is recovered.Note: Your final reports in

Page 31 - Key Label Keyboard key

Quinton® Q-Stress® 4.5 70-00755-01 D10-110 SetupThis chapter describes the steps in setting up the parameters for the exercise device, protocol and p

Page 32 - Keyboard Equivalents

70-00755-01 D Quinton® Q-Stress® 4.510-2SetupDemonstration ModeThe system ships with a demonstration mode that provides you with a way to train techn

Page 33 - Software

Quinton® Q-Stress® 4.5 70-00755-01 D10-3Setting up the ProcedureSetting up the ProcedureA procedure defines those parameters used in a particular str

Page 34 - Q-Stress Application

70-00755-01 D Quinton® Q-Stress® 4.510-4SetupECG TabUse this section to select Lead Set, Filters, Worst Case leads, and ST Measurement Point (Figure

Page 35 - User Documentation

70-00755-01 D Quinton® Q-Stress® 4.51-2SafetyIndications for Use◆ The device is intended to acquire, process, record, archive, analyze, and output el

Page 36 - 4 System Setup

Quinton® Q-Stress® 4.5 70-00755-01 D10-5Setting up the ProcedureNotification TabThis section (Figure 10-4) allows you to set a base point for visual

Page 37 - System Setup

70-00755-01 D Quinton® Q-Stress® 4.510-6SetupFormulas TabThis section (Figure 10-5) allows you to set formulas for METS Baseline, Target Heart Rate,

Page 38 - Set up an email account

Quinton® Q-Stress® 4.5 70-00755-01 D10-7Setting up the ProcedurePrompts for Ramped ProtocolsFigure 10-7: Example of Ramped Protocol promptsIndicate p

Page 39 - Configure Adobe Acrobat

70-00755-01 D Quinton® Q-Stress® 4.510-8SetupPrompts for Staged ProtocolsFigure 10-8: Example of Staged Protocol promptsStaged Protocol prompts are s

Page 40 - Electrodes

Quinton® Q-Stress® 4.5 70-00755-01 D10-9Setting up the ProcedureScreen Layout TabThis section (Figure 10-9) allows you to select the positioning of v

Page 41 - Connection Test

70-00755-01 D Quinton® Q-Stress® 4.510-10SetupLead FormatClick the radio button in front of 3 Lead, 6 Lead or 12 lead to select the default format.Cl

Page 42 - Display Options

Quinton® Q-Stress® 4.5 70-00755-01 D10-11Setting up the Procedurespace. After you have cleared a space, you can introduce it again by clicking on the

Page 43 - Skin Preparation

70-00755-01 D Quinton® Q-Stress® 4.510-12SetupRPEUse this box to enter the rate of perceived exertion. Click the down arrow to display the numerical

Page 44

Quinton® Q-Stress® 4.5 70-00755-01 D10-13Setting up the ProcedureSpO2This box allows the entry of a patient’s blood oxygen saturation level. If Q-Str

Page 45 - Quik-Prep Tips

70-00755-01 D Quinton® Q-Stress® 4.510-14SetupTimed ReportsWhen you click the Timed Reports tab, the system will always display the Staged Timed Repo

Page 46

Quinton® Q-Stress® 4.5 70-00755-01 D1-3Safety DefinitionsSafety DefinitionsThe following types of safety statements are used in this manual.!WARNINGI

Page 47 - Lead configurations

Quinton® Q-Stress® 4.5 70-00755-01 D10-15Setting up the ProcedureRamped Timed ReportsEnter the time and double-click in the appropriate column to sch

Page 48

70-00755-01 D Quinton® Q-Stress® 4.510-16SetupTo create a custom protocol:1. From the main menu, select Setup | Protocol (Ctrl+P). The Protocol Setup

Page 49

Quinton® Q-Stress® 4.5 70-00755-01 D10-17Setting up the Procedure3. Select a protocol close to the one you want to create. It will be shown in the Pr

Page 50 - Entering Patient Information

70-00755-01 D Quinton® Q-Stress® 4.510-18SetupErgometer ProtocolAdd customized ergometer protocols using the same directions as given for creating a

Page 51 - Adding a Patient Entry

Quinton® Q-Stress® 4.5 70-00755-01 D10-19Set Up SystemSet Up SystemVarious lists, devices and the measurement unit used in the test must be set up be

Page 52 - Previous Studies/Tests

70-00755-01 D Quinton® Q-Stress® 4.510-20SetupRest labels, Comments, Personnel, Institution, Locations, Resting ECG, and Billing Codes lists are ship

Page 53 - Target HR

Quinton® Q-Stress® 4.5 70-00755-01 D10-21Set Up SystemChanging Personnel ListsTo add a new entry or change an existing entry, click the Personnel rad

Page 54

70-00755-01 D Quinton® Q-Stress® 4.510-22SetupChanges to these lists are made the same way. Click the list you would like to change. The list is disp

Page 55 - Protocol (required)

Quinton® Q-Stress® 4.5 70-00755-01 D10-23Set Up SystemThe Location list is the drop-down list for location on the Environment tab in patient entry (F

Page 56 - Final Report (required)

70-00755-01 D Quinton® Q-Stress® 4.510-24SetupTo edit an entry, triple-click the item. This will take you to the edit mode and allow you to place the

Page 57 - Medications (Optional)

70-00755-01 D Quinton® Q-Stress® 4.51-4SafetyGeneral Safety StatementsThis section lists general safety statements for the Q-Stress system. Additiona

Page 58

Quinton® Q-Stress® 4.5 70-00755-01 D10-25Set Up SystemDevice TabClick this tab to assign specific devices to the COM ports.Figure 10-28: Example of t

Page 59 - Diagnosis (Optional)

70-00755-01 D Quinton® Q-Stress® 4.510-26SetupPrinter SelectionClick this tab (Figure 10-29) to configure the Q-Stress program for the printer attach

Page 60

Quinton® Q-Stress® 4.5 70-00755-01 D10-27Set Up SystemPDF Location (Network Option Only)Change the location where PDF files are stored either by dire

Page 61 - Risk Factors (System Option)

70-00755-01 D Quinton® Q-Stress® 4.510-28SetupQ-Exchange® Tab (Q-Exchange Option Only)This tab (Figure 10-31) allows you to set the version number of

Page 62

Quinton® Q-Stress® 4.5 70-00755-01 D11-111 OptionsChapter contents◆ Custom Reports 11-2◆ Q-Exchange® Patient Data Export and Import 11-6◆ Full Disclo

Page 63 - Print Patient Info

70-00755-01 D Quinton® Q-Stress® 4.511-2OptionsCustom ReportsThe custom report template enables summary reports to be formatted with the information

Page 64 - 7 Running a Stress Test

Quinton® Q-Stress® 4.5 70-00755-01 D11-3Custom Reports4. In the dialog box (Figure 11-2), click OK to start InfoPath. You may receive a warning from

Page 65

70-00755-01 D Quinton® Q-Stress® 4.511-4OptionsNote: Here are some tips for working with the Infopath Designer:To place a field onto the canvas, drag

Page 66 - Blood Pressure Monitor

Quinton® Q-Stress® 4.5 70-00755-01 D11-5Custom ReportsImporting Custom Pages into Q-Stress1. After editing and saving the form, click Import in the F

Page 67 - Running a Stress Test

70-00755-01 D Quinton® Q-Stress® 4.511-6OptionsQ-Exchange® Patient Data Export and ImportQ-Exchange 4.0 can export all data gathered during a stress

Page 68

Quinton® Q-Stress® 4.5 70-00755-01 D1-5Symbols and LabelsSymbols and LabelsThe following symbols may be used in this manual, related documentation, o

Page 69

Quinton® Q-Stress® 4.5 70-00755-01 D11-7Q-Exchange® Patient Data Export and Import◆ Full disclosure—Unfiltered, un-transformed waveform binary data a

Page 70

70-00755-01 D Quinton® Q-Stress® 4.511-8Options*_RE.xml—Report ECG in XML format*_SR.xml—Stress record in XML format; this file is written regardless

Page 71

Quinton® Q-Stress® 4.5 70-00755-01 D11-9Full DisclosureFull DisclosureOn Q-Stress systems equipped with the Full Disclosure option, a continuous log

Page 72 - Close the Patient Entry

70-00755-01 D Quinton® Q-Stress® 4.511-10OptionsPage Attachment BarSelect/Unselect Current Page The Select/Unselect Current Page button toggles the s

Page 73 - Data and Display Options

Quinton® Q-Stress® 4.5 70-00755-01 D11-11Full DisclosureList of PagesThe List of Pages gives navigation cues and attachment status of the Full Disclo

Page 74

70-00755-01 D Quinton® Q-Stress® 4.511-12OptionsNavigation BarThe Navigation Bar allows you to quickly display pages in the Displayed Beats portion o

Page 75 - Cancel Test

Quinton® Q-Stress® 4.5 70-00755-01 D11-13ReanalysisReanalysisThe Reanalysis option provides you with a mechanism to review a complete test after the

Page 76 - Change ASVV

70-00755-01 D Quinton® Q-Stress® 4.511-14OptionsThe Reanalysis window opens with the current average beat set to the worst case average beat for the

Page 77 - Change Ergometer Workload

Quinton® Q-Stress® 4.5 70-00755-01 D11-15Reanalysisoriginal positions as determined by the original procedure. An edited J+ point affects all average

Page 78 - Change Displayed Leads

70-00755-01 D Quinton® Q-Stress® 4.511-16OptionsZoom In/OutThe Zoom button toggles your view between normal and maximum size. Click Zoom to change th

Page 79 - Change Protocol

70-00755-01 D Quinton® Q-Stress® 4.51-6SafetyAHA to IEC EquivalentsThis manual and the Q-Stress software use the AHA label (RL, LL, etc.) in diagrams

Page 80 - Left Decrease speed

Quinton® Q-Stress® 4.5 70-00755-01 D11-17ReanalysisClick a single graph to zoom in on it (Figure 11-12).Figure 11-12: Example of an expanded graphCli

Page 81 - Comments

70-00755-01 D Quinton® Q-Stress® 4.511-18OptionsFreeze FrameThe Freeze Frame option allows total recall of ECGs during a test. You can “freeze” a 10-

Page 82 - Display Trend Graphs

Quinton® Q-Stress® 4.5 70-00755-01 D11-19Freeze Framepane. Freeze Frame continues adding data to the right of the pane, refreshing approximately ever

Page 83 - ER Graph

70-00755-01 D Quinton® Q-Stress® 4.511-20OptionsSizing the Freeze Frame Pane (4)To ensure that the real time data is not obscured, the Freeze Frame p

Page 84 - SpO2 Graph

Quinton® Q-Stress® 4.5 70-00755-01 D11-21Risk ScoringRisk ScoringThe system includes a risk scoring option based on the Duke, Framingham, and Functio

Page 85 - Ectopic Rate

70-00755-01 D Quinton® Q-Stress® 4.511-22OptionsEntering Risk ParametersMost of the parameters used for risk scoring are entered with the patient dem

Page 86 - Exit the Application

Quinton® Q-Stress® 4.5 70-00755-01 D11-23Risk ScoringDuke Treadmill (TM) The Duke TM algorithm requires the Angina index to calculate risk. The Angin

Page 87

70-00755-01 D Quinton® Q-Stress® 4.511-24OptionsFraminghamSource: Wilson PWF, D'Agostino RB, Levy D, Belanger AM, Silbershatz H, Kannel WB. Pred

Page 88

Quinton® Q-Stress® 4.5 70-00755-01 D11-25Risk ScoringThe algorithm uses the Total Cholesterol Score Sheet when total cholesterol and HDL cholesterol

Page 89

70-00755-01 D Quinton® Q-Stress® 4.511-26OptionsThe Duke treadmill score is calculated as follows: Treadmill score = duration of exercise in minutes

Page 90

Quinton® Q-Stress® 4.5 70-00755-01 D2-12 Quick StartThis section lists the basic steps required to perform a stress test. If necessary, refer to othe

Page 91 - Hold Stage

Quinton® Q-Stress® 4.5 70-00755-01 D11-27Risk ScoringEnter any or all of the Risk Scoring data into the final report by double-clicking on the value

Page 92 - Print ECGs During Test

70-00755-01 D Quinton® Q-Stress® 4.511-28OptionsAutomatic Blood Saturation Level (SpO2) MonitoringQ-Stress provides an interface for SpO2 data to be

Page 93 - Print Ectopic Beat Report

Quinton® Q-Stress® 4.5 70-00755-01 D11-29Ramped Protocol GeneratorRamped Protocol GeneratorWith this option installed, Q-Stress can generate custom r

Page 94 - QRS Sync Lead

70-00755-01 D Quinton® Q-Stress® 4.511-30Options4. Press Ta b to move between fields or click a field to select it. You can type in a number or use

Page 95

Quinton® Q-Stress® 4.5 70-00755-01 D11-31Ramped Protocol Generator• For a treadmill, type the target METs, maximum speed, and the duration of the Exe

Page 96 - Rest Label

70-00755-01 D Quinton® Q-Stress® 4.511-32Options

Page 97

Quinton® Q-Stress® 4.5 70-00755-01 D12-112 TroubleshootingChapter contents◆ Software Troubleshooting 12-2◆ User Messages/Errors 12-3◆ Service Tests 1

Page 98 - ST Average Beat

70-00755-01 D Quinton® Q-Stress® 4.512-2TroubleshootingSoftware TroubleshootingLanguage Settings for Adobe ReaderUse the following steps if Adobe Rea

Page 99 - Start/Stop Belt

Quinton® Q-Stress® 4.5 70-00755-01 D12-3User Messages/ErrorsUser Messages/ErrorsUser Message Sys. Log Message Cause ActionBPBP monitor data error.BP

Page 100 - 8 Final Reports

70-00755-01 D Quinton® Q-Stress® 4.512-4TroubleshootingTreadmill motor over heated.Treadmill motor over heated.Treadmill motor over heated.Allow 10 s

Page 101 - Final Reports

70-00755-01 D Quinton® Q-Stress® 4.52-2Quick StartBefore the TestPerform these steps to prepare for the test.1. Explain the procedure to the patient.

Page 102 - (Figure 8-2)

Quinton® Q-Stress® 4.5 70-00755-01 D12-5User Messages/ErrorsElectrodesElectrode [name] Disconnected.Leads Off.Electrode lead has disconnected from pa

Page 103 - Fast Reports

70-00755-01 D Quinton® Q-Stress® 4.512-6TroubleshootingThe preamp unit is reporting a self test error. Contact Technical Support.ECG card reported se

Page 104 - Review Mode

Quinton® Q-Stress® 4.5 70-00755-01 D12-7User Messages/ErrorsBlanking of ECG data.Your PC system can overload. If an over- loading of the system occur

Page 105 - Print a Final Report

70-00755-01 D Quinton® Q-Stress® 4.512-8TroubleshootingTCR-1000 Chart Recorder TroubleshootingYour chart recorder is designed and manufactured to pro

Page 106 - Saving the Test

Quinton® Q-Stress® 4.5 70-00755-01 D12-9Hewlett-Packard (HP) LaserJet Printer TroubleshootingHewlett-Packard (HP) LaserJet Printer TroubleshootingIn

Page 107 - Lead Set

70-00755-01 D Quinton® Q-Stress® 4.512-10TroubleshootingSQL ErrorIf a message is displayed indicating the SQL server is not running, use the followin

Page 108

Quinton® Q-Stress® 4.5 70-00755-01 D12-11Service TestsService TestsAfter installing or upgrading the software, run the following tests to ensure the

Page 109 - Averages

70-00755-01 D Quinton® Q-Stress® 4.512-12Troubleshooting2. Click Start Self Test. The program runs the self-test and the following dialog box is disp

Page 110 - Peak Exercise

Quinton® Q-Stress® 4.5 70-00755-01 D12-13Service Tests3. Click Next. The Gain and Impedance results appear4. Review Table 12-1 on page 12-15 to see i

Page 111

70-00755-01 D Quinton® Q-Stress® 4.512-14TroubleshootingLeads Off TestThe leads-off test checks the preamp’s leads-off detection capability. This tes

Page 112 - In-Test Reports

USER’S GUIDE AND SERVICE MANUALQUINTON® Q-STRESS®VERSION 4.570-00755-01 D

Page 113

Quinton® Q-Stress® 4.5 70-00755-01 D2-3Rest PhaseRest PhaseThe Rest phase should not begin until after the patient is prepared for testing. During th

Page 114 - 9 File Management

Quinton® Q-Stress® 4.5 70-00755-01 D12-15Service Tests8. Click Back to return to the test or Close to end the test.Service ReportTo view a Service Re

Page 115 - Database

70-00755-01 D Quinton® Q-Stress® 4.512-16Troubleshooting

Page 116 - Final Report Repository

Quinton® Q-Stress® 4.5 70-00755-01 D13-113 MaintenancePerform the routine maintenance procedures in this chapter per the schedule listed or as needed

Page 117 - PDF Files

70-00755-01 D Quinton® Q-Stress® 4.513-2MaintenanceOnce per ShiftThe following procedures must be performed once per shift, daily, or as needed.!WARN

Page 118 - Archive PDF files

Quinton® Q-Stress® 4.5 70-00755-01 D13-3Annual◆ Cords: Ensure all power cords and other cables are not worn or damaged. Specifically, check for loose

Page 119 - Restore PDF Files

70-00755-01 D Quinton® Q-Stress® 4.513-4MaintenanceComponent ReplacementNormally, if a a board or other part of component is damaged or fails, the en

Page 120 - Database Backup

Quinton® Q-Stress® 4.5 70-00755-01 DA-1A SpecificationsAppendix contents◆ System Specifications A-2◆ Ventilation Clearances A-6◆ Fuse Replacement A-6

Page 121 - File Management

70-00755-01 D Quinton® Q-Stress® 4.5A-2System SpecificationsSystem specifications are listed belowTable A-1: System Specifications OverallPerformance

Page 122 - Database Restore

Quinton® Q-Stress® 4.5 70-00755-01 DA-3System SpecificationsInput Channels Simultaneous acquisition of all leadsChart Recorder (optional)Type Multipl

Page 123 - Purge the Database

70-00755-01 D Quinton® Q-Stress® 4.5A-4Type 19 in. LCD flat panelMemory SpanMinimum of 8 seconds of ECG (with two average beats also displayed)Displa

Page 124

70-00755-01 D Quinton® Q-Stress® 4.52-4Quick StartWarm-up PhaseAfter you have taken the resting ECG and entered other Rest data, have the patient war

Page 125

Quinton® Q-Stress® 4.5 70-00755-01 DA-5System SpecificationsPatient CableDomestic Cable10-electrode with pinch or snap connectors, blue, 15 foot (4.6

Page 126 - 10 Setup

70-00755-01 D Quinton® Q-Stress® 4.5A-6Ventilation ClearancesAllow at least 6 inches behind each workstation for ventilation. There are no side clear

Page 127 - Demonstration Mode

Quinton® Q-Stress® 4.5 70-00755-01 DA-7Q-Stress Customer-supplied Computer SpecificationsQ-Stress Customer-supplied Computer SpecificationsSpecificat

Page 128 - Setting up the Procedure

70-00755-01 D Quinton® Q-Stress® 4.5A-8Uninterruptible Power Supply (UPS) RecommendationsThe Q-Stress system requires a steady supply of power during

Page 129 - Worst Case

Quinton® Q-Stress® 4.5 70-00755-01 DA-9Uninterruptible Power Supply (UPS) RecommendationsTable A-2: UPS for Chart Recorder Systems 110 VACFunction Sp

Page 130 - Notification Tab

70-00755-01 D Quinton® Q-Stress® 4.5A-10Table A-4: UPS for Laser Printer Systems 110 VACFunction SpecificationMaximum Load 1400 VA / 950 WInput Volta

Page 131 - Prompts Tab

Quinton® Q-Stress® 4.5 70-00755-01 DA-11EMC Guidance and Manufacturer’s DeclarationEMC Guidance and Manufacturer’s DeclarationThe Q-Stress System is

Page 132 - Prompts for Ramped Protocols

70-00755-01 D Quinton® Q-Stress® 4.5A-12Stress System as recommended below, according to the maximum output power of the communications equipment.Ele

Page 133 - Prompts for Staged Protocols

Quinton® Q-Stress® 4.5 70-00755-01 DA-13EMC Guidance and Manufacturer’s DeclarationSurgeIEC 61000-4-5± 1 kV differential mode±2 kV common mode± 1 kV

Page 134 - Screen Layout Tab

70-00755-01 D Quinton® Q-Stress® 4.5A-14Radiated RFIEC 61000-4-33 V/m80 MHz to 2.5 GHz3 V/mc 80 MHz to 800 MHz 800 MHz to 2.5 GHzwhere P is the maxim

Page 135 - Info Toolbar Tab

Quinton® Q-Stress® 4.5 70-00755-01 D2-5Recovery PhaseRecovery PhaseWhen the patient has completed the Exercise phase (or for some reason must discont

Page 136

Quinton® Q-Stress® 4.5 70-00755-01 DA-15Treadmill/ErgometerTreadmill/ErgometerSee treadmill or ergometer documentation for specifications.

Page 137

70-00755-01 D Quinton® Q-Stress® 4.5A-16

Page 138 - Report Format Tab

Quinton® Q-Stress® 4.5 70-00755-01 DB-1B Protocol ParametersThe tables in this appendix detail the exercise protocols stored in Q-Stress. These inclu

Page 139 - Timed Reports

70-00755-01 D Quinton® Q-Stress® 4.5B-2System-Supplied ProtocolsBruce ProtocolStage No. Duration Speed (mph) Grade (%)Rest/Recovery -- 1.2 0.01 3:00

Page 140 - Setting up Protocols

Quinton® Q-Stress® 4.5 70-00755-01 DB-3System-Supplied ProtocolsNaughton Protocol Stage No. Duration Speed (mph) Grade (%)Rest/Recovery -- 1.0 0.01 2

Page 141

70-00755-01 D Quinton® Q-Stress® 4.5B-4Modified Balke Protocol Stage No. Duration Speed (mph) Grade (%)Rest/Recovery -- 1.2 0.011:003.4 0.021:003.4 1

Page 142

Quinton® Q-Stress® 4.5 70-00755-01 DB-5System-Supplied ProtocolsRamped Low Protocol Time Stage NoSpeed (mph) Grade (%) METsRest/Recov 1.0 0.0 1.80:30

Page 143 - Pharmacological Protocol

70-00755-01 D Quinton® Q-Stress® 4.5B-6Ramped Medium Protocol Time Stage NoSpeed (mph) Grade (%) METsRest/Recov 1.0 0.0 1.80:30 1 1.5 3.0 2.80:30 2 1

Page 144 - Set Up System

Quinton® Q-Stress® 4.5 70-00755-01 DB-7System-Supplied ProtocolsRamped High Protocol Time Stage NoSpeed (mph) Grade (%) METsRest/Recov 1.0 0.0 1.80:3

Page 145

70-00755-01 D Quinton® Q-Stress® 4.5B-8Åstrand Protocol (Ergometer)Stage No. Duration Workload (Watts)Rest/Recovery -- 5016:00502 6:00 1003 6:00 1504

Page 146 - Changing Comment Type Lists

70-00755-01 D Quinton® Q-Stress® 4.52-6Quick Start

Page 147

Quinton® Q-Stress® 4.5 70-00755-01 DC-1C Taking Accurate Blood Pressure MeasurementsAppendix contents◆ BP412 and STBP-780 C-2◆ Attaching the Cuff C-3

Page 148 - Changing Medication Lists

70-00755-01 D Quinton® Q-Stress® 4.5C-2BP412 and STBP-780Note: The BP412 (or the STBP-780 for Japanese units only) meet the AAMI automated blood pres

Page 149

Quinton® Q-Stress® 4.5 70-00755-01 DC-3BP412 and STBP-780Use these settings for the blood pressure (BP) device:◆ Set Systolic Max to 300◆ Set Systoli

Page 150 - Device Tab

70-00755-01 D Quinton® Q-Stress® 4.5C-44. Ensure that the bottom of the cuff is at least one inch above the bend in the elbow (Figure C-3).Figure C-3

Page 151 - Printer Selection

Quinton® Q-Stress® 4.5 70-00755-01 DC-5BP412 and STBP-780these readings and uses this information when taking the more noisy exercise measurements.◆

Page 152

70-00755-01 D Quinton® Q-Stress® 4.5C-6SunTech® Tango+® Blood Pressure MonitorThe following instructions show you how to attach the SunTech Tango+ Bl

Page 153

Quinton® Q-Stress® 4.5 70-00755-01 DC-7SunTech® Tango+® Blood Pressure MonitorPlugging in the RS-232 cableTo plug in the RS-232 cable:1. Plug the RS-

Page 154 - 11 Options

70-00755-01 D Quinton® Q-Stress® 4.5C-8Setting up a Blood Pressure MonitorTo configure a blood pressure monitor on a Stress system:1. On the desktop,

Page 155 - Custom Reports

Quinton® Q-Stress® 4.5 70-00755-01 DD-1D Manual Calibration of Cardiac Science TreadmillsTM and ST Series treadmills do not require calibration at in

Page 156 - Modify the report template

70-00755-01 D Quinton® Q-Stress® 4.5D-2

Page 157 - Saving the Custom Design

Quinton® Q-Stress® 4.5 70-00755-01 D3-13 System OverviewThis chapter lists the standard Q-Stress system components and features. For details refer to

Page 158

Quinton® Q-Stress® 4.5 70-00755-01 DE-1E Glossary and AbbreviationsAppendix contents◆ Glossary E-2◆ Abbreviations E-7

Page 159 - Exporting data

70-00755-01 D Quinton® Q-Stress® 4.5E-2GlossaryTerm Definition12-lead ReportA recording of sequential or simultaneous segments of the conventional le

Page 160

Quinton® Q-Stress® 4.5 70-00755-01 DE-3GlossaryECG Complex One complete cycle of an ECG waveform (see next figure). The baseline provides the referen

Page 161 - Importing data

70-00755-01 D Quinton® Q-Stress® 4.5E-4In-test ReportReports printed manually or automatically during any phase of a stress test. Examples include 12

Page 162 - Full Disclosure

Quinton® Q-Stress® 4.5 70-00755-01 DE-5GlossaryMax HR x BPEach time a BP is entered (either manually or automatically), the systolic value is multipl

Page 163 - Page Attachment Bar

70-00755-01 D Quinton® Q-Stress® 4.5E-6Sequential 12-Lead A 12-lead report that prints the data from three or six leads synchronously for the specifi

Page 164 - List of Pages

Quinton® Q-Stress® 4.5 70-00755-01 DE-7AbbreviationsAbbreviationsASVV Absolute Spatial Vector VelocityBP Blood Pressure, expressed as Systolic/Diast

Page 165 - Navigation Bar

70-00755-01 D Quinton® Q-Stress® 4.5E-8

Page 166 - Reanalysis

70-00755-01 D*70-00755-01*Cardiac Science Corporation • 3303 Monte Villa Parkway, Bothell, WA 98021 USA • 425.402.2000 • US toll-free 800.426.0337 • F

Page 167 - Changing Measurement Points

70-00755-01 D Quinton® Q-Stress® 4.53-2System Overview!WARNING. Patient injury.Q-Stress supports the use of Trackmaster treadmill model TMX425 as an

Page 168 - Measurement Point Toolbar

Quinton® Q-Stress® 4.5 70-00755-01 D3-3HardwareHardware!WARNING. Patient or operator injury.Ensure that components of the Stress system are placed in

Page 169 - Trend Graph

70-00755-01 D Quinton® Q-Stress® 4.53-4System OverviewPort Use1AC Power2115 or 230 V.AC Selector. Must be set to match incoming voltage.3AC Power Swi

Page 170

Quinton® Q-Stress® 4.5 70-00755-01 D3-5HardwarePreamp IIThe Preamp II digitizes and filters patient signals. A data collector is integrated with the

Page 171 - Freeze Frame

70-00755-01 D Quinton® Q-Stress® 4.53-6System OverviewThe isolation transformer is plugged into a dedicated circuit. The following components are plu

Page 172 - Vital Statistics (3)

Information in this document is subject to change without notice. Names and data used in the examples are fictitious unless otherwise noted.CE Mark De

Page 173 - Closing the Pane (6)

Quinton® Q-Stress® 4.5 70-00755-01 D3-7HardwareKeypadThe optional keypad (Figure 3-2) controls the treadmill or ergometer and the phases of the Q-Str

Page 174 - Risk Scoring

70-00755-01 D Quinton® Q-Stress® 4.53-8System OverviewAdditional system functions may be accessed using key combinations. The following table lists t

Page 175 - Entering Risk Parameters

Quinton® Q-Stress® 4.5 70-00755-01 D3-9HardwareKeyboard EquivalentsNon-English versions of Q-Stress may have non translated icons or keys. The follow

Page 176

70-00755-01 D Quinton® Q-Stress® 4.53-10System OverviewSoftwareThe Q-Stress system computer is pre-loaded with the Q-Stress application and the Micro

Page 177 - Framingham

Quinton® Q-Stress® 4.5 70-00755-01 D3-11SoftwareSystem ShutdownTo shut down the Q-Stress system on Windows 7 and Vista:1. Select File | Exit from the

Page 178 - Duke Treadmill

70-00755-01 D Quinton® Q-Stress® 4.53-12System OverviewFigure 3-3: Example of a test windowUser DocumentationThis manual is available from the Help m

Page 179 - Viewing Risk Scoring Results

Quinton® Q-Stress® 4.5 70-00755-01 D4-14 System SetupThis section lists the basic software setup and configuration procedures to install and properly

Page 180

70-00755-01 D Quinton® Q-Stress® 4.54-2System SetupInstall Virus Scanning SoftwareRunning virus scanning software while Q-Stress is in use can cause

Page 181

Configure the Operating SystemThe Windows operating system is pre-configured for use with Q-Stress.!WARNING. Inaccurate test data.Changing any BIOS pa

Page 182 - Ramped Protocol Generator

70-00755-01 D Quinton® Q-Stress® 4.54-4System SetupConfigure Adobe AcrobatThe Q-Stress system includes Adobe® Reader® to view final reports as PDF fi

Page 183 - For an Individual Patient

70-00755-01 D ContentsiiiContentsContentsChapter 1: SafetyIndications for Use...

Page 184 - Removing a Custom Protocol

Quinton® Q-Stress® 4.5 70-00755-01 D5-15 Connecting and Checking the ElectrodesCardiac Science recommends the Quik-Prep™ stress test electrode applic

Page 185

70-00755-01 D Quinton® Q-Stress® 4.55-2Connecting and Checking the ElectrodesConnection TestBefore you begin the Rest phase of a stress test, you mus

Page 186 - 12 Troubleshooting

Quinton® Q-Stress® 4.5 70-00755-01 D5-3Connection Test5. Click Start to begin the Impedance Check. The message Checking lead set, please wait appears

Page 187 - Software Troubleshooting

70-00755-01 D Quinton® Q-Stress® 4.55-4Connecting and Checking the ElectrodesSkin PreparationA good, low-impedance electrical connection between the

Page 188 - User Messages/Errors

Quinton® Q-Stress® 4.5 70-00755-01 D5-5Skin Preparation6. Disconnect the recharger from the DC jack.7. Plug the reference wire into the RL (R) jack,

Page 189 - Troubleshooting

70-00755-01 D Quinton® Q-Stress® 4.55-6Connecting and Checking the Electrodes!WARNING. Skin trauma.Due to physiological differences in electrode site

Page 190 - 70-00755-01 D

Quinton® Q-Stress® 4.5 70-00755-01 D5-7Skin Preparation2. Check the quality of the right leg electrode preparation.If the right leg site is inadequat

Page 191

70-00755-01 D Quinton® Q-Stress® 4.55-8Connecting and Checking the ElectrodesLead configurationsThe following figures indicate where to place the ele

Page 192

Quinton® Q-Stress® 4.5 70-00755-01 D5-9Skin PreparationTable 5-3: Frank Lead Placement SitesSite LWN1Patient Cable Color Frank Lead Placement SitesEV

Page 193

70-00755-01 D Quinton® Q-Stress® 4.55-10Connecting and Checking the Electrodes

Page 194

Contents 70-00755-01 DivContentsChapter 4: System SetupInstall Virus Scanning Software...4-

Page 195 - SQL Error

Quinton® Q-Stress® 4.5 70-00755-01 D6-16 Entering Patient InformationQ-Stress allows you to store patient information and stress test data in a datab

Page 196 - Service Tests

70-00755-01 D Quinton® Q-Stress® 4.56-2Entering Patient InformationAdding a Patient EntryClick the New Patient Test icon. The Patient Selection scree

Page 197 - Performance Test

Quinton® Q-Stress® 4.5 70-00755-01 D6-3Adding a Patient Entry1Patient Demographic Information2Test Parameters3Previous Studies/Tests{{{123

Page 198 - Click Close to end this test

70-00755-01 D Quinton® Q-Stress® 4.56-4Entering Patient InformationPatient Demographic InformationType in the demographics information required by yo

Page 199 - Leads Off Test

Quinton® Q-Stress® 4.5 70-00755-01 D6-5Test ParametersTest ParametersThere are seven tabbed sections used to set parameters: Environment, Order Info,

Page 200 - Service Report

70-00755-01 D Quinton® Q-Stress® 4.56-6Entering Patient InformationExercise Device (required)Select the exercise device the patient will be using for

Page 201

Quinton® Q-Stress® 4.5 70-00755-01 D6-7Test ParametersProcedure (required)The procedure is a set of parameters that defines how an exercise test will

Page 202 - 13 Maintenance

70-00755-01 D Quinton® Q-Stress® 4.56-8Entering Patient InformationOrder Info (Optional)Use the Order Info window (Figure 6-3) to enter transaction i

Page 203 - Once per Shift

Quinton® Q-Stress® 4.5 70-00755-01 D6-9Test ParametersUse this window (Figure 6-5) to enter medications that the patient is currently taking. The sys

Page 204 - Clean and Disinfect

70-00755-01 D Quinton® Q-Stress® 4.56-10Entering Patient Informationarrows to move the month forward and back. Click the year to display arrows to ch

Page 205 - Component Replacement

70-00755-01 D ContentsvContentsChapter 7: Running a Stress TestBefore the Test...

Page 206 - A Specifications

Quinton® Q-Stress® 4.5 70-00755-01 D6-11Test ParametersClinical Procedures (Optional)Use the Clinical Proc tab (Figure 6-7) to enter other clinical p

Page 207 - System Specifications

70-00755-01 D Quinton® Q-Stress® 4.56-12Entering Patient InformationResting ECGs (Optional)Use the Resting ECGs tab (Figure 6-8) to record the “botto

Page 208

Quinton® Q-Stress® 4.5 70-00755-01 D6-13Previous Studies/TestsPrevious Studies/TestsThe lower portion of the Patient Information window shows previou

Page 209 - ), user-selectable

70-00755-01 D Quinton® Q-Stress® 4.56-14Entering Patient InformationPrint Patient InfoClick this button for a printout of the patient’s information.

Page 210

Quinton® Q-Stress® 4.5 70-00755-01 D7-17 Running a Stress TestThe test has four phases: Rest, Exercise, Recovery, and Review. The instructions below

Page 211 - BTU Output

70-00755-01 D Quinton® Q-Stress® 4.57-2Running a Stress TestBefore the TestThe system limits the total time in Rest, Exercise, and Recovery phases to

Page 212

Quinton® Q-Stress® 4.5 70-00755-01 D7-3Blood Pressure MonitorBlood Pressure MonitorIf you have an automatic blood pressure device attached to your st

Page 213 - Recommendations

70-00755-01 D Quinton® Q-Stress® 4.57-4Running a Stress TestBlood Saturation Level (SpO2) ReportingBlood saturation levels can be entered manually du

Page 214 - Function Specification

Quinton® Q-Stress® 4.5 70-00755-01 D7-5Rest PhaseRest PhaseThe Rest phase begins after you have prepared the patient for testing. During this phase,

Page 215

70-00755-01 D Quinton® Q-Stress® 4.57-6Running a Stress Testthe change to be processed. Once the Start Exercise icon is active again, you can advance

Page 216 - Electromagnetic Emissions

Contents 70-00755-01 DviContentsPrint ECGs During Test...7-29Print Ectopic Beat Rep

Page 217 - Electronic Immunity

Quinton® Q-Stress® 4.5 70-00755-01 D7-7Warm-up PhaseWarm-up PhaseAfter you have performed the resting ECG and gathered other rest data, have the pati

Page 218

70-00755-01 D Quinton® Q-Stress® 4.57-8Running a Stress TestWhen a notification event occurs, a message appears. The message stays until you remove i

Page 219

Quinton® Q-Stress® 4.5 70-00755-01 D7-9Initial Editing of the Final Report Summary PageInitial Editing of the Final Report Summary PageDuring the Rec

Page 220 - Treadmill/Ergometer

70-00755-01 D Quinton® Q-Stress® 4.57-10Running a Stress TestData and Display OptionsDuring a test, you can enter data or change the way ECG channels

Page 221

Quinton® Q-Stress® 4.5 70-00755-01 D7-11Data and Display OptionsRefer to the blood pressure device user guide for the for more information. See Ta k

Page 222 - B Protocol Parameters

70-00755-01 D Quinton® Q-Stress® 4.57-12Running a Stress TestCancel TestSee Abort or Cancel Test on page 7-10.

Page 223 - System-Supplied Protocols

Quinton® Q-Stress® 4.5 70-00755-01 D7-13Change ASVVChange ASVVSelect Tools | ASVV Selection. The ASVV Lead Selection dialog box appears (Figure 7-2).

Page 224

70-00755-01 D Quinton® Q-Stress® 4.57-14Running a Stress TestChange Ergometer WorkloadClick Load. The Up and Down arrows appear. Click the Up arrow t

Page 225 - Modified Balke Protocol

Quinton® Q-Stress® 4.5 70-00755-01 D7-15Change Displayed LeadsChange Displayed LeadsTo change the leads displayed, click the lead label you want to c

Page 226 - (mph) Grade (%) METs

70-00755-01 D Quinton® Q-Stress® 4.57-16Running a Stress TestChange ProtocolTo change the protocol during the test, click the word Protocol. A drop-d

Page 227 - Ramped Medium Protocol

70-00755-01 D ContentsviiContentsChapter 10: SetupDemonstration Mode ...10-2Set

Page 228 - Ramped High Protocol

Quinton® Q-Stress® 4.5 70-00755-01 D7-17Change Treadmill Speed or GradeChange Treadmill Speed or GradeAvailable: Rest, Exercise, and Recovery phases.

Page 229 - Stage No. Duration

70-00755-01 D Quinton® Q-Stress® 4.57-18Running a Stress TestCommentsAvailable: Rest, Exercise, and Recovery phasesTo enter comments during the test,

Page 230 - Measurements

Quinton® Q-Stress® 4.5 70-00755-01 D7-19Display Trend GraphsDisplay Trend GraphsAvailable: Exercise and Recovery phasesTo display a trend graph durin

Page 231 - BP412 and STBP-780

70-00755-01 D Quinton® Q-Stress® 4.57-20Running a Stress TestHR GraphThe HR graph (Figure 7-5) shows Heart Rate in beats per minute over time.Figure

Page 232 - Attaching the Cuff

Quinton® Q-Stress® 4.5 70-00755-01 D7-21Display Trend GraphsST Level or ST Slope GraphsThe ST Level or Slope graphs (Figure 7-8) show (by individual

Page 233 - Microphones

70-00755-01 D Quinton® Q-Stress® 4.57-22Running a Stress TestEctopic RateAvailable: Rest, Exercise, and Recovery phases.This automatic function calcu

Page 234 - Patient Instructions

Quinton® Q-Stress® 4.5 70-00755-01 D7-23Exit the ApplicationExit the ApplicationAfter you complete the test, you can exit Q-Stress by selecting File

Page 235 - Requirements

70-00755-01 D Quinton® Q-Stress® 4.57-24Running a Stress TestFiltersAvailable: Rest, Exercise, and Recovery phasesYou can activate various filters to

Page 236 - Setting up the Tango+ Monitor

Quinton® Q-Stress® 4.5 70-00755-01 D7-25Freeze Frame (Freeze Frame Option Only)Freeze Frame (Freeze Frame Option Only)See Freeze Frame on page 11-18

Page 237

70-00755-01 D Quinton® Q-Stress® 4.57-26Running a Stress TestGainAvailable: Rest, Exercise, and Recovery phasesTo change the gain for the waveforms,

Page 238 - Science Treadmills

Contents 70-00755-01 DviiiContentsFreeze Frame...11-18Initiating Fr

Page 239

Quinton® Q-Stress® 4.5 70-00755-01 D7-27GradeGradeSee Change Treadmill Speed or Grade on page 7-17.

Page 240 - E Glossary and Abbreviations

70-00755-01 D Quinton® Q-Stress® 4.57-28Running a Stress TestHold StageSee Advance to Next Exercise Stage on page 7-10.

Page 241 - Glossary

Quinton® Q-Stress® 4.5 70-00755-01 D7-29Print ECGs During TestPrint ECGs During TestAvailable in Rest, Exercise and Recovery phasesNote: Do not turn

Page 242 - Q-Stress uses

70-00755-01 D Quinton® Q-Stress® 4.57-30Running a Stress TestPrint Ectopic Beat ReportAvailable: Rest, Exercise, and Recovery phasesThe Ectopic Beat

Page 243 - ST Level Sample Point

Quinton® Q-Stress® 4.5 70-00755-01 D7-31QRS Sync LeadQRS Sync LeadAvailable: Rest, Exercise, and Recovery phases The default QRS sync lead is V5. Occ

Page 244 - Term Definition

70-00755-01 D Quinton® Q-Stress® 4.57-32Running a Stress TestRate of Perceived Exertion (RPE)Available: Rest, Exercise, and Recovery phasesIf this bu

Page 245

Quinton® Q-Stress® 4.5 70-00755-01 D7-33Rest LabelRest LabelAvailable: Rest phaseThe Rest Label indicates the patient’s position or physiological sta

Page 246 - Abbreviations

70-00755-01 D Quinton® Q-Stress® 4.57-34Running a Stress TestSpeedSee Change Treadmill Speed or Grade on page 7-17.

Page 247

Quinton® Q-Stress® 4.5 70-00755-01 D7-35ST Average BeatST Average BeatAvailable: Overall Worst Case in Exercise and Recovery phases; Current Worst Ca

Page 248 - *70-00755-01*

70-00755-01 D Quinton® Q-Stress® 4.57-36Running a Stress TestStart/Stop BeltAvailable: Rest, Exercise, and Recovery phasesThis button toggles between